Balance in the open airflow
It is also a common disappointment. The air in the bench moves at 0.45 m/s to keep the work clean.
This moving air is enough to disturb an analytical balance at the resolution most laboratories weigh to.
Fabtech supplies laminar airflow workbenches that give ISO Class 5 air over the work. A LAF workbench is a work bench with clean air moving over it.
It is also called a LAF bench, a clean bench or a LAF work station. All these names mean the same machine.
It is a work surface with filtered air moving across it in one direction. The work stays in this clean air even when the room air is less clean.
We reply within 24 hours.
Our GMP and FDA-compliant LAF workbenches give product protection for small to medium work processes at laboratory scale. The protection comes from directed airflow.
Clean air with low turbulence (very little swirling) flows onto the work area. The air then goes back into the room at table height.
A sheet metal cassette element limits the airflow at the back. Polycarbonate panes limit it at the sides.
The motor blower is statically and dynamically balanced. Because of this, it keeps the capacity and static pressure needed for the whole life of the filter.
The bench collects airborne particles and discharges them in a controlled way.
Key fact Detail
Fabtech supplies one machine under all four names.
They mean the same type of equipment. Different markets use different names for it.
A clean bench is a LAF bench. A LAF work station is a LAF workbench.
The name does not change the machine.
The units differ in these ways:
Airflow direction. Vertical airflow moves down onto the work.
Horizontal airflow moves from the back of the bench toward you.
Enclosure. The front can be open, or it can have sliding panes.
The panes can also have glove ports, for more separation.
Scale. This page covers bench-scale units.
A larger process area needs a vertical LAF unit.
A clean bench is not a biological safety cabinet.
The air moves from the filter, across your work, and then out toward you.
This protects the sample. It gives the operator no protection at all.
Do not use a clean bench for anything infectious, cytotoxic (toxic to living cells) or potent. For this work you need a Class II biological safety cabinet, or a containment isolator at negative pressure.
This is the most common misuse of a clean bench in a working laboratory.

Weighing inside a clean bench is a common request.
It is also a common disappointment. The air in the bench moves at 0.45 m/s to keep the work clean.
This moving air is enough to disturb an analytical balance at the resolution most laboratories weigh to.
Weighing system integration is available, and it works. The layout must shelter the balance from the direct airflow.
It must also avoid a stagnant pocket, which is a spot where the air does not move.
Tell us at enquiry stage if a balance is going in. The balance changes the internal design, so we design it in from the start.
It is not added afterwards.
For cleanroom equipment for projects in India, contact Fabtech Technologies Cleanrooms Ltd, a separate and separately listed group company.
It is a workstation that gives a clean, particle-free work zone. Filters remove contaminants from the incoming air. The filtered air then passes over the work area in one direction, as a uniform stream.
There is no difference. They are different names for the same equipment. Units differ in airflow direction, enclosure and size.
Fabtech supplies it in powder-coated SS 304 or SS 316. Both are grades of stainless steel.
It keeps contamination to a minimum and holds sterile conditions. Because of this, the ordinary air movement of the room does not put critical work at risk.
Pharmaceuticals, biotechnology, life sciences, healthcare, and food and beverage. Fabtech supplies them for pharma, biopharma and healthcare projects.
Laminar air moves in one direction at a uniform velocity. Turbulent air moves in several directions at once. This mixed movement is what carries particles onto a work surface.
No. It protects the product only. The air flows across the work and out toward the operator. For operator protection you need a biological safety cabinet or a containment isolator.
Place it in the designated cleanroom zone. Turn it so the airflow direction suits the space and the process. Leave clearance around it for maintenance. Provide a stable electrical supply. Plan for full DQ/IQ/OQ/PQ validation, the formal tests that prove the unit works as required.
Yes, with weighing system integration, if the layout shelters the balance from direct airflow. Tell us at enquiry stage, not after installation.
Tell us your bench size and what you handle on it, and we will specify the unit
We reply within 24 hours.